Quadrus MINI Imager User’s Manual
13-19
Utilities
Customer Default Parameters
Customer default parameters (saved by
<Zc>
) are the same set of parameters as power-on
parameters but are saved in a different, isolated section of NOVRAM. This allows a user
essentially to create a backup set of parameters that can be recalled in the event that the
current parameters or power-on parameters have been accidentally changed or are no
longer desired.
It is important to note that a hardware default does not affect customer default parameters.
For example, a user that has inadvertently changed communication settings and saved
them with a
<Z>
command, may not know the correct settings or have the capability to
communicate within those settings. By first doing an EZ button default or hardware default
to restore the known Microscan defaults, the user can then recall the previous customer-
saved settings with an
<Arc>
or
<Zrc>
command.
Microscan Default Parameters
Microscan default parameters are contained in the firmware and cannot be changed.
Software Defaults
Microscan default parameters can be recalled (loaded into current settings) with
<Ard>
command or recalled and saved for power-on with the
<Zrd>
command.
Hardware Default
If a software default and reset is not possible, it may be necessary to reset the reader by
shorting (connecting) specific pins. This procedure has the same effect as the
<Zrd>
software command.
Important: For this reset to occur, this command must be executed within 60 seconds of a
power-on or a reset.
1. Apply power to the reader.
2. If using an IB-131, locate pins 7 and 11 on the host connector.
Caution: Be certain that the correct pins are located. Connecting the wrong pins could
cause serious damage to the unit.
3. Momentarily connect these wires (or pins) and listen for a series of short beeps.
4. Within 3 seconds, connect them again. A longer beep should be heard. If not, repeat
the process.
Default on Power-On
You can also use the EZ button to default the reader by holding it down while applying
power to the reader, provided that this feature is enabled.
Summary of Contents for Quadrus MINI
Page 1: ...Quadrus MINI Imager User s Manual P N 83 006300 Rev H...
Page 12: ...xii Quadrus MINI Imager User s Manual Statement of RoHS Compliance...
Page 98: ...4 20 Quadrus MINI Imager User s Manual Additional Notes about Calibration...
Page 128: ...5 30 Quadrus MINI Imager User s Manual Setting Up the Imager for EZ Trax...
Page 180: ...6 52 Quadrus MINI Imager User s Manual Background Color...
Page 238: ...7 58 Quadrus MINI Imager User s Manual Configuring EZ Trax Output...
Page 298: ...10 32 Quadrus MINI Imager User s Manual Skew Correction...
Page 318: ...11 20 Quadrus MINI Imager User s Manual Request All Configuration Database Settings...
Page 348: ...13 24 Quadrus MINI Imager User s Manual Other Operational Serial Commands...
Page 362: ...14 14 Quadrus MINI Imager User s Manual Ordered Output Filter...