U N I T S
L E V E L :
m m
D I S P L A Y :
%
T E M P E R A T U R E
°C
D A M P I N G
0 5
UNITS
Level meter can process and convert a large number of different
physical values
. The setting
is done in the item "UNITS".
LEVEL: Unit selection (mm, cm, m, in, ft)
DISPLAY: The unit showed on the display
(%, mm, cm, m, in, ft, l, hl, m
3
, gal, bbl,
mA)
TEMPERATURE: Temperature unit (°C, °F)
DAMPING
Setting the
response time
of the measurements. The function is useful for suppressing
level fluctuations, waves and rapid changes of the level.
The reaction time will depend on the
exponential function. Damping with a defined delay in seconds represents the time when
exponential reaches 2/3 of its maximum value.
The damping time can be set in the
interval from 0 to 99 s.
1. To enter to the menu press
the same button to select "BASIC SETTINGS". Then, using
and
select "UNITS".
2.
Now the menu item "UNITS" is shown.
By pressing the
and
button make the set
-
tings of individual items.
3.
By pressing
button save the data. By next presses of the button
leave the menu.
The level meter returns to measurement mode.
1. To enter to the menu press
the same button to select "BASIC SETTINGS". Then, using
and
select "DAMPING".
2.
Now the menu item "DAMPING" is shown.
By pressing the
and
button make the set
-
tings of individual items.
3.
By pressing
button save the data. By next presses of the button
leave the menu.
The level meter returns to measurement mode.
1. To enter to the menu press
the same button to select "BASIC SETTINGS".
Then, using
and
select
"MIN LEVEL" or "MAX LEVEL".
2.
Now it is shown the item "MIN LEVEL" ("MAX LEVEL"). By pressing
and
set the out-
put current
"OUTPUT", the distance for the defined current "LEVEL" the value on the dis
-
play
"DISPLAY".
3.
By pressing
button save the data. By next presses of the button
leave the menu.
The level meter returns to measurement mode.
M I N L E V E L
M A X L E V E L
U N I T S
D A M P I N G
S E N S I T I V I T Y
T E A C H I N G
M I N L E V E L
M A X L E V E L
U N I T S
D A M P I N G
S E N S I T I V I T Y
T E A C H I N G