Dinel, s.r.o.      ULM–70

   

You can set the sensitivity in three 

degrees: 

LOW – MEDIUM – HIGH. 

SENSITIVITY

The setting is defined in three steps of the level meter 

sensitivity

"LOW" 

–  Low sensitivity in case of surrounding interferences affecting the measurement. 

"MEDIUM"  – 

Medium sensitivity (suitable for most applications).

"HIGH"  –  Enhanced sensitivity for measured media partly absorbing the ultrasonic signal 

If there are no above obstacles 

in the tank, it is not necessary to 

start this mode.

TEACHING

The mode serves for 

suppressing false reflections

 resulting from reflection of the ultra

-

sonic signal from roughnesses on walls of the tank, various partitions, mixers or other obsta

-

cles. The sensor starting this mode detects false reflections and save them in the memory. 

Then these false reflections will not affect the subsequent measurement (they are masked).

Before starting the mode it is necessary to empty the tank as much as possible (pref-

erably completely).

1.  To enter to the menu press

the same button to select "SERVICE". Then, using

and

select "TEACHING".

2. 

Now it is shown the item "TEACHING". By pressing

set the value "LEVEL DISTANCE" (dis

-

tance to the level) – supposed distance from the face of the sensor to the medium level. If the 

distance to the level is not precisely known, enter a value rather lower (in the tolerance field 

as shown in Fig. 19).

3. 

After entering the "SET LEVEL DISTANCE" by pressing

button the system starts "teaching" 

(false reflection mapping). During the mapping, the display shows flashing sign "RUNNING".

4. 

The mapping of false echoes can be completed when you see the inscription "Press OK to 

stop" and you press

.

5. 

The procedure is completely finished when you can see the inscription "DONE". It is then 

possible to exit the menu by repeated pressing the button

.

S E N S I T I V I T Y

M E D I U M

T E A C H I N G

S E T   L E V E L   D I S T A N C E :

0 6 0 0 0   m m

        S T A R T

M I N   L E V E L

M A X   L E V E L

U N I T S

D A M P I N G

S E N S I T I V I T Y

T E A C H I N G

M I N   L E V E L

M A X   L E V E L

U N I T S

D A M P I N G

S E N S I T I V I T Y

T E A C H I N G

T E A C H I N G

S E T   L E V E L   D I S T A N C E :

0 6 0 0 0   m m

        R U N N I N G

T E A C H I N G

S E T   L E V E L   D I S T A N C E :

0 6 0 0 0   m m

    

D O N E

P R E S S

 

E S C

  T O   E X I T

The mode "TEACHING" will stop 

automatically after ca. 1000 meas

-

urements.

T E A C H I N G

S E T   L E V E L   D I S T A N C E :

0 6 0 0 0   m m

        R U N N I N G

P R E S S

 

O K

  T O  

S T O P

If during the scanning of the tank in the bottom of the display 

appears the dialog "press OK to stop" (see figure) the level me

-

ter already found no further obstacles and "TEACHING" mode 

may be terminated. If it is not terminated, the level meter is still 

ready for the possible presence of obstacles (e.g. paddles of 

the agitator). Once it registers a further obstacle, the dialogue 

disappears and the obstacle is erased. This process may be 

repeated up to 1000 cycles. After this the "TEACHING" mode is 

automatically stopped.

Summary of Contents for Dinel ULM-70

Page 1: ...ished in this manual before the first use of the level meter Keep the manual at a safe place The manufacturer reserves the right to implement changes without prior notice INSTRUCTION MANUAL ultrasonic...

Page 2: ......

Page 3: ...he level meter with current output and MGU unit 13 8 4 Wiring diagram of the level meter with Modbus RS485 and MGU unit 13 8 5 Wiring diagram of the level meter with Modbus RS485 and PDU unit 13 9 Set...

Page 4: ...vessels sumps reservoirs and open channels In case the level of bulk solid materials is measured the meas urement range is reduced The level meters can continuously measure levels of bulk solid ma ter...

Page 5: ...70_ 02 Pg11 cable gland Ground terminal variant ULM 70 with protective conductor ULM 70_ 06 ULM 70_ 20 ULM 70_ 10 This procedure includes the following three steps INSTALLATION ELECTRIC CONNECTION SE...

Page 6: ...essary to install the level meter so that the bin level cannot interfere with the dead zone when filled up to the maximum If the measured level in terferes with the dead zone the level meter will not...

Page 7: ...meter installation outside the influence of filling In case the level of bulk solids is measured the measurement range is reduced Due to absorption of acoustic waves by a bulk me dium shortening of th...

Page 8: ...level meter Fig 9 For such cases it is necessary to set up SENSITIV ITY mode to high or contact the manufacturer if need In case of a thin layer of foam it is also possible to use the directional hor...

Page 9: ...o prevent penetration of humidity Then the rain and condensing water can flow off freely The cable bushing and connector have to be sufficiently tightened to prevent penetration of humidity To lower t...

Page 10: ...e level meter with Modbus is involved as a terminal for RS 485 we recommend to avoid reflections on the line to connect 120 termi nation resistor This is done by moving a small lever of the switch mar...

Page 11: ...14 15 R1 3 4 optional ACTIVE current output R2 n c n c optional optional Power supply depending on version 1 2 24V 10 11 12 13 14 15 3 4 OC1 OC2 OC1 OC4 Umax 30V DC Imax 30mA Pmax 100mW OC3 OC4 option...

Page 12: ...10 Imax 100 mA PE n c PDU 420 W 230V input 4 20 mA Connection of PDU 420 W is valid for firmware version 6 00 or higher The older ver sions up to version 5 99 the terminal 9 of the IRU unit is connect...

Page 13: ...als description OC type outputs 11 GND DATA DATA RS 485 Uo 24V 5 10 Imax 100mA 1 2 5 6 7 8 9 10 11 12 13 14 15 3 4 31 32 33 34 35 OC1 OC2 OC1 OC4 Umax 30V DC Imax 30mA Pmax 100mW OC3 OC4 optional opti...

Page 14: ...ND MGU UNIT 12 13 14 18 19 A B _ S RS 485 Master 24V DC Imax 120 mA A YE B GN 0V WH U BN Cab Sh ULM 70N _ _ _ M A YE B GN 0V WH U BN Cab Sh MGU 800 without additional modules 8 4 WIRING DIAGRAM OF THE...

Page 15: ...ntly TEACHING mode activation Symbol E 1 Lighting intermittently correct echo receiving of the reflected signal from the measured surface level Symbol 1 Lighting permanently level meter is locked agai...

Page 16: ...ITS If in the bottom of the display appears when entering the values the inscription OUT OF LIM ITS the value specified for the item LEVEL is outside the measuring range of the level meter If the insc...

Page 17: ...ssing the and button make the set tings of individual items 3 By pressing button save the data By next presses of the button leave the menu The level meter returns to measurement mode 1 To enter to th...

Page 18: ...wn enter a value rather lower in the tolerance field as shown in Fig 19 3 After entering the SET LEVEL DISTANCE by pressing button the system starts teaching false reflection mapping During the mappin...

Page 19: ...AR T FA CTORY DE FAULT R ESET MEDIUM TEMPERATURE YES 20 C Inactive compensation initial state the word NO appears on the display See the UNITS menu for tempera ture unit selection C or F After start o...

Page 20: ...LT ARE YOU SURE FACTORY DEFAULT RUNNING FACTORY DEFAULT DONE PRESS ESC TO EXIT RESET ARE YOU SURE During the restart process RUN NING will be displayed Then the level meter will be automatically turne...

Page 21: ...ng The CLONE SETTINGS mode transfers all data excluding setting of the Teaching and HART BASIC SETTINGS SERVICE DIAGNOSTIC CLONE SETTINGS PASSWORD INFO CLONE SETTINGS NOW CLONING CL ONE SE TTING S SEN...

Page 22: ...e words NO PASSWORD will be displayed 2 Now you can edit the password The actual edited item is displayed inversely Press the but ton to move to the next position clockwise direction button serves to...

Page 23: ...2 Protocol HART Universalcommunicationinterfacefordatacommunicationofperipheraldeviceswiththelevel meter Data transmission runs through the same line as the 4 20mA current loop without impact on analo...

Page 24: ...onductor Set up elements D basic version with display L without display metal lid Performance N Normal usable in non explosive areas only Xi Ex proof suitable for explosive areas Maximum reach 15 Acce...

Page 25: ...al conditions for safe use ULM 70Xi The device is designed for connection to the isolating repeater IRU 420 When the other ap proved supply unit is used whose output parameters satisfy above mentioned...

Page 26: ...t 30 70 C a II 1 2 G Ex ia IIB T5 Ga Gb P 0 99 W I 132 mA i i C 370 nF L 0 9 mH i i FTZ 09 ATEX 0277X 1026 I 4 20 mA ULM 70Xi 06 G I B D IP67 1026 ULM 70Xi 10 G I B D IP67 1026 ULM 70Xi 20 F I B D IP6...

Page 27: ...B D ULM 70N __ _ M _ _ Varianty zna en 01 02 06 10 20 D L B H G F Labels for type of ULM 70N _ _ _ M _ _ Example of label for type of ULM 70N 10 G M B D Symbol of producer logo Dinel Internet address...

Page 28: ...T T I N G S S E R V I C E D I A G N O S T I C C L O N E S E T T I N G S P A S S W O R D L A N G U A G E I N F O M I N L E V E L M A X L E V E L U N I T S D A M P I N G S E N S I T I V I T Y T E A C H...

Page 29: ...e ULM 70_ 02 06 ULM 70_ 10 20 30 70 C 30 60 C Short time temperature stress resistance 90 C 1 hour Max operation overpressure on transmission surface 0 1 MPa Sensitivity 3 steps low medium high Dampin...

Page 30: ...module Display type Matrix OLED Resolution 128 x 64 pixel Character height Number of digits measured value 9 mm 5 Digits Display colour Yellow Buttons Membrane switch panel Ambient temperature range 3...

Page 31: ...l indicates particularly important characteristics of the device Note This symbol indicates helpful additional information 21 Packaging shipping and storage The device DLM 35 is packaged in a polyethy...

Page 32: ...002 003 e mail sale dinel cz www dinel cz The lastest version of this instruction manual can be found at www dinel cz Version 06 2019 This project was realised under financial support from the funds o...

Reviews: