Page 31
Outputs and Inputs
thermoMETER CTVideo
8.
Outputs and Inputs
8.1
Analog Output
This output is used for the object temperature. The selection of the output signal can be done via the pro-
gramming keys, see Chap.
9.
The software allows the programming of output channel 1 as an alarm output.
Output signal
Range
Connection pin on CTVideo board
Voltage
0 ... 5 V
OUT-mV/mA
Voltage
0 ... 10 V
OUT-mV/mA
Current
0 ... 20 mA
OUT-mV/mA
Current
4 ... 20 mA
OUT-mV/mA
Thermocouple
TCJ
OUT-TC
Thermocouple
TCK
OUT-TC
i
According to the chosen output signal there are different connection pins on the mainboard
(OUT-mV/mA or OUT-TC).
8.2
Digital Interface
Die thermoMETER CTVideo sensor is equipped with
an USB interface. The interface board is located
beside the LCD display to uninstall the board please
disconnect both of the M3x5 screws.
i
Please pay attention to the correct positioning of
the pin strip if you install the board.
Fig. 13 Connection board for the USB interface