
Page 41
Operation
thermoMETER CTVideo
9.5
Focusing and Video Sighting
On the back plane of the sensor you will find
a rotary button for focusing of the optics.
To set the focus to the desired
measurement distance connect the
sensor with a PC using the USB cable
and start the software.
You will see the video beside the tem-
perature-time diagram. The position of the
measurement spot is indicated by a circle
inside the video picture. The size of this
circle is equivalent to the IR spot size.
By turning the focusing rotary button, see
Fig. 17
, clockwise you will change the focus
in direction far.
By turning counterclockwise you will change
the focus in direction close.
After a successful focusing please en-
ter the measurement distance (distance
sensor front – object) in the according
field in the software (underneath the
video picture, see
Fig. 18
).
You will find a detailed description of the vid-
eo settings in the software description which
you can call via the menu
? > Help....
Focusing rotary button
Fig. 17 Focusing by the focusing rotary button
Measurement distance
Fig. 18 View video presentation
Summary of Contents for thermoMETER CTVideo
Page 6: ...thermoMETER CTVideo...
Page 63: ......