Page 34
Operation
thermoMETER CTVideo
9.
Operation
After power up the unit the sensor starts an initializing routine for some seconds. During this time the display
will show
INIT
. After this procedure the object temperature is shown in the display. The display backlight
color changes according to the alarm settings, see Chap.
8.4
.
9.1
Restoring Factory Setting
To set the thermoMETER CTVideo back to the factory default settings, please press at first the
button
and then the
button and keep both pressed for approx. 3 seconds.
The display will show
RESET
for confirmation.
Display
Modus (Example)
Adjustment Range
S ON
Laser Sighting (On)
ON/OFF
142.3C
Object temperature (after signal processing)
(142.3 °C)
Fixed
127CH
Sensor temperature (127 °C)
Fixed
25CB
Box temperature (25 °C)
Fixed
142CA
Actual object temperature (142 °C)
Fixed
MV5
Signal output channel 1 (0 - 5 V)
0 - 20 = 0 - 20 mA/
4 - 20 = 4 - 20 mA/
MV5 = 0 - 5 V/
MV10 = 0 - 10 V/
TCJ = Thermo couple type J/
TCK = Thermo couple type K
E0.970
Emissivity (0,970)
0,100 ... 1,100
T1.000
Transmissivity (1,000)
0,100 ... 1,100
A 0.2
Signal output Average (0.2 s)
A---- = inactive/0,1 ... 999,9 s
P----
Signal output Peak hold (inactive)
P---- = inactive/0,1 ... 999,9 s/P
∞
∞
∞
∞
=
infinite