Page 27
Installation
optoNCDT 22xx
Cable Requirement
5.5
Power supply:
Controller 2200:
24 VDC (±15 %, max. 500 mA)
-
screened cable, screen connected with the plugbody
-
Connect the screen of the power cable with the safety earth conductor.
Voltage output:
Max. length 10 m (32 ft), the electromagnetic field may cause measurement uncertainty on the signal if you
work with cables longer then 10 m (32 ft). MICRO-EPSILON recommends to terminate the end of the cable
with 10 nF to avoid noise voltages.
Twisted wires
-
Screened cable, screen connected with the plug body
-
Connect the screen with the safety earth conductor
Error output and synchronization:
Twisted wires
-
Screened cable, screen connected with the plug body
-
i
When using power supply units, always use devices which are VDE-conform and tested!
For controller with integrated power supply unit:
- Only use a power cable with a protective circuit connection (inlet connector for non-heating appli-
ances)!
- Connection to a socket outlet with an earthing contact required!