Appendix | ASCII Communication with Sensor
optoNCDT 1220
A 3.3.4.5 ERRORLEVELOUT1, Output Level Digital Output
ERRORLEVELOUT1 NPN|PNP|PUSHPULL|PUSHPULLNEG
Choice of output level for ERROROUT1.
- NPN: digital output is active in case of an error
- PNP: digital output is active in case of an error
- PUSHPULL: digital output with high level in case of an error
- PUSHPULLNEG: digital output with low level in case of an error
Wiring of digital output ERROR1, see Chap.
A 3.3.4.6 ERRORLIMIT
ERRORLIMIT DIST1 <upper threshold>
Digital output is activated, if the defined value for a measurement is exceeded.
Range: -2 ... 2 * measuring range [mm].
A 3.3.4.7 ERRORHYSTERESIS
ERRORHYSTERESIS <hysteresis>
Digital output is deactivated, if the defined value for a measurement falls below the limit value.
Range: 0 ... 2 * measuring range [mm].
A 3.3.4.8 ERROROUTHOLD
ERROROUTHOLD <hold period>
Indicates in ms how long the digital output must be active at least when the limit value is exceeded. This time
period starts when the limit value is exceeded. Range: 0 ... 1000 [ms].