Ethernet Interface
6.
Ethernet Interface
You will achieve especially high resolutions if you readout the measurements in digital
form via the Ethernet interface.
For that purpose, use the web interface or the enclosed runtime version on CD-ROM,
see Chap.
6.5
, or use your own program. Micro-Epsilon supports the MEDAQLib driver
that includes all commands for the capaNCDT 6500. You will find more details on the
enclosed CD or on the Internet: www.micro-epsilon.com/en/software “Standard applica-
tions > MEDAQlib“.
6.1
Hardware, Interface
In order to use the Ethernet interface, a demodulator module must be provided in chan-
nel 1 as this one determines the cycle for all channels!
The data logging of all channels is synchronous.
Connect the DT6530 to an available Ethernet interface at the PC. Use a crossover
cable.
For a connection with the DT6530 you will require a defined IP address of the network
interface card inside the PC. Go to Control Panel\Network. Set up, if applicable, a new
LAN connection. For more information, contact you network administrator.
Fig. 32 LAN-Connection of a PC
Define the following address in the properties of the LAN connection:
IP
address:
169.254.168.1
Subnet mask: 255.255.0.0
Select