Technical Data
Controller type
DT6530
DL6530
DL6510 with CP6001
DT6530
DL6510 mit CPM6011
Resolution, static
(2.6 Hz)
0.000075 % FSO
0.0006 % FSO
Resolution,
dynamic
(100 Hz)
0.0003 % FSO
0.0025 % FSO
(1 kHz)
0.0009 % FSO
0.007 % FSO
(8.5 kHz)
0.002 % FSO
0.015 % FSO
Limit frequency analog output
20 Hz; 1 kHz; 8.5 kHz (-3 dB, adjustable)
Data rate output
Ethernet
3.9 kSa/s
7.8 kSa/s (max. 4 channels)
EtherCAT
2.0 kSa/s
Linearity (typical)
±0.025 % FSO
±0.05 % FSO
Maximum sensitivity deviation
±0.05 % FSO
±0.1 % FSO
Repeatability
1
0.0003 % FSO
0.001 % FSO
Long time stability
±0.002 % FSO / month
±0.02 % FSO / month
Synchronous operation
2
yes
yes
Insulator measurement
yes
no
Temperature stability
digital: 5 ppm/°C
analog: 10 ppm/°C
80 ppm
(digital and analog)
Operating temperature, sensor
-50 ... +200 °C (-58 ... +392 °F)
Operating temperature,
controller
+10 ... +60 °C (+50 ...+140 °F)
Storage temperature
-10 ... +75 °C (+14 ... +167 °F)
Power supply
100 ... 240 VAC (50 ... 60 Hz)
optional: 18 ... 36 VDC
Output
0 ... 10 V (max. 10 mA, short circuit proof);
offset
≤
±10 V up to 0 V
4 ... 20 mA (max. load 500 Ohm)
optional: 0 ... 20 mA (max. load 500 Ohm)
Ethernet 24 bit / EtherCAT 24 bit
Sensors
all sensors are suitable
Sensor cable length, standard
CCm1.4x
CCg2.0x
Sensor cable length, matched
≤ 4.2 m (type CCmx)
≤ 8.0 m (type CCgx)
≤ 2.8 m (type CCmx)
≤ 4 m (type CCgx)
Trigger
TTL, 5 V
Humidity
5 ... 95 % (non-condensing)
Protection class
IP 40 (controller and sensors)
FSO = Full Scale Output
1) at constant ambient temperature (including temperature and air humidity)
2) Possible to further Controller DT6530 resp. DT6530C