Operation
capaNCDT 6500
Page 27
Trimmer Zero:
Turn the trim-pot clockwise to shift the characteristic line to the left.
Zero
EMR
0
1
Signal
Displacement
EMR = End of measuring range
i
The Zero-Poti affects only the analog outputs, not yet the digital measuring values.
Target selection
Use the slide switch, see
Fig. 27
, to switch between conducting (
Cond
.) and non-con-
ducting targets (
Insul
.). In position
Cond
. (conductor) only the zero setting with the ze-
ro-trimmer is active. The amplifying is set to 0 up to 10 V at the entire measuring range.
Insul.
Cond.
Fg0
Fg1
Fg2
Fig. 27 Switches on board for material selection and limit frequency
Limit frequency analog output
The limit frequency of the analog output signal can be set with a rotary switch on the
board, see
Fig. 27
.
Three positions are possible:
Limit frequency fg0 = 8.5 kHz
Limit frequency fg1 = 1 kHz
Limit frequency fg2 = 20 Hz