XTP501 & XTC501 User’s Manual
14
99976 Issue 1.1, May 2020
OPERATION
2.5.3 Secondary Parameters Page
PCB TEMP, °C 28
CELL T, °C 50.0
COMP I/P
OFF
EXT I/P
OFF
Figure 12
Secondary Parameters Page
Parameter Description
CELL T
Sensor cell temperature display in set unit (°C, °F or Kelvin)
Display Resolution = 0.1
PCB TEMP
Temperature display of Microcontroller in selected temperature unit
This gives an indication of the internal temperature
Display Resolution = 1 unit
Accuracy = ±2ºC
COMP I/P
The value of compensation input (mA i/p channel 1) as a % (4 mA=0%
and 20 mA=100%)
OFF displayed instead of value if external compensation is turned off
EXT I/P
Value of the external input in the selected parameter and unit
(DEWP, TEMPR, PRESS or NONE)
OFF displayed instead of value if external compensation parameter is set
to NONE