XTP501 & XTC501 User’s Manual
Michell Instruments
13
OPERATION
2.5.2 Chart Page
t
t-4
25
%
0
Figure 11
Chart Page
NOTE : This data is not available via the Modbus
•
This indicative chart is continuously running at the set chart interval
(2 to 60 seconds).
•
Chart duration in seconds = (chart interval * 60).
•
It is auto ranging with an auto range resolution of 1%.
•
It is reset if the chart interval is changed or the instrument power is
cycled.
•
Chart data is only stored in volatile memory and therefore is not saved.
•
Chart interval is saved and available in a Modbus register.
•
Chart data is not available via serial comms as the Application Software is
able to perform more sophisticated charting functions