XTP501 & XTC501 User’s Manual
Michell Instruments
15
OPERATION
2.5.4 Min/Max Page
MINIMUM 0.00
%CO2
D12/01 T 19:29:44
D12/01 T 19:29:44
MAXIMUM 0.00
%CO2
Figure 13
Min/Max Page
This indicates the minimum and maximum values measured, along with date/time of
occurrence. The value is reset manually via the Reset Page in the User Menu.
NOTE: This data is not saved in non-volatile (NV) memory and is not available
via serial communications.
2.5.5 Alarms Log Pages
Figure 14
Alarms Log Page
A maximum of 40 High/Low alarms, along with date and time of occurrence, are
recorded in a ring buffer in NV memory. The most recent alarm will overwrite the oldest
alarm when more than 40 alarms are recorded. The data is displayed over a maximum
of 10 pages (with 4 alarms on each page). The latest alarm record is displayed in line
1 of page 1. This data is not available via serial communications or in the blind unit.
The data is reset manually via the Reset Page in the User Menu. The data is saved and
restored when the instrument is restarted. P1 = Page 1.