Test results, shown in the QA-1290's digital display, can be printed
out directly, or transferred to a PC via the
ansur
QA-1290 test auto-
mation software.
ansur
lets you design test protocols, remotely con-
trol the QA-1290, and store the test results.
1.2
Specifications
1.
Simulations
Simulation technique:
Oscillometric
Calibration Tables used: DINAMAP, HP-Merlin
Simulation repeatability:
±0.5 mmHg
Heart Rate:
10-200 bpm, 1 bpm steps
Heart Rate Accuracy:
±1 bpm @pp< 2 mmHg
Pulse Pressure:
0.05-5.00 mmHg
Systolic/Diastolic Pres-
sure:
As shown below
Preset Pressures
(Systolic/Diastolic (MAP), in mmHg)
Adult
Neonate
60/ 30 (40)
60/ 30 (40)
80/48 (58)
80/48 (58)
100/65 (77)
100/65 (77)
120/80 (95)
120/80 (95)
150/95 (114)
150/95 (114)
200/140 (167)
200/140 (167)
255/195 (215)
255/190 (220)
Independent Systolic/Diastolic Settings
(in mmHg)
Systolic
20 to 255
Diastolic
5 to 220
Pulse Pressure
0.05 to 5.00
Preset Artifacts and Arrhythmias
Artifacts include: Bradycardia, Tachycardia,
Geriatric, Obese, Strenuous Exercise, Mild
Exercise, and Weak Pulse
Arrhythmias include: Premature Atrial Con-
traction, Premature Ventricular Contraction,
Atrial Fibrillation, and 2
nd
Degree Heart Block
Custom Patient Pattern Mode
1-2
Summary of Contents for QA-1290
Page 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Page 6: ...This page intentionally left blank vi...
Page 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Page 30: ...This page intentionally left blank 4 8...
Page 36: ...This page intentionally left blank 5 6...
Page 51: ...Appendix A Tubing Kit Components A 1...
Page 52: ...This page intentionally left blank B 2...
Page 54: ...B 2...
Page 55: ...B 3...
Page 56: ...B 4...
Page 57: ...B 5...
Page 58: ...B 6...
Page 59: ...B 7...
Page 60: ...B 8...
Page 61: ...B 9...