2. Press
SET UP (F3)
. This advances you to the Inflate Pressure
Screen.
(See below).
Manometer: 0 mmHg ZERO PR.> F7
PRINT...> F6
Set Pressure: 300 mmHg ...............> F5
REGULATE SET UP MAIN MENU
F1 F2 F3 F4
Test object: Monitor ................> F7
F6
F5
STORE GO BACK MAIN MENU
F1 F2 F3 F4
3. Press
Test object (F7)
. This allows you to select between
Moni-
tor
and
Cuff and Monitor
as the port being used
.
4. Press
STORE (F1)
to save the settings then press
GO BACK
(F2)
to return to the manometer test.
Manometer: 0 mmHg ZERO PR.> F7
PRINT...> F6
Set Pressure: 300 mmHg ..............> F5
REGULATE SET UP MAIN MENU
F1 F2 F3 F4
5. To reset the manometer to 0 mmHg, press
ZERO PR (F7)
.
6. If you want to print the test results, press
PRINT (F6)
to advance
to the Print Screen.
7. Press
Set Pressure (F5)
to key in the pressure that you wish the
QA-1290 to maintain.
8. Once you have completed your setup press
REGULATE (F2)
to
begin pressure regulation.
Note:
It is possible to change the pressure units for the Leak,
Manometer and OverPressure tests. From the Main Menu press
(F3) setup and more (F1) 3 times until the menu item is dis-
played.
Pressure Unit: mmHg………………………> F7
Press F7 for the desired pressure units. mmHg, cmH20, inH20,
PSI, kPa, or mBar.
5-5
Summary of Contents for QA-1290
Page 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Page 6: ...This page intentionally left blank vi...
Page 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Page 30: ...This page intentionally left blank 4 8...
Page 36: ...This page intentionally left blank 5 6...
Page 51: ...Appendix A Tubing Kit Components A 1...
Page 52: ...This page intentionally left blank B 2...
Page 54: ...B 2...
Page 55: ...B 3...
Page 56: ...B 4...
Page 57: ...B 5...
Page 58: ...B 6...
Page 59: ...B 7...
Page 60: ...B 8...
Page 61: ...B 9...