At the end of the transfer, a normal startup sequence will occur.
4.
Test to confirm the upgrade.
•
Exit the DOS program and unhook the RS-232 transfer ca-
ble from the QA-1290.
•
Switch the QA-1290
OFF
, then turn it back
ON
.
Verify that the new firmware version is installed while the QA-
1290 is running its start sequence. The firmware version will be
displayed in Startup Screen Menu 3.
3-8