2.
Press
SETUP (F3)
. This advances you to the Start Pressure
Screen.
(See below).
Start Pressure: XXX mmHg ......> F7
Test object : Cuff ......> F6
F5
STORE GO BACK MAIN MENU
F1 F2 F3 F4
3. Use the alphanumeric keys to enter the
Start Pressure
(F7)
set-
ting to between 0 and 400 mmHg. Press the
Enter
key to save
the new setting(s). Use the
Delete
key to clear the current setting.
4. Scroll through
Test Object (F6)
to enter the device under test.
Available options are: Cuff, Monitor, and Cuff and Monitor.
5. Press:
•
STORE (F1)
to store the settings in flash memory.
•
GO BACK (F2)
to return to the Leak Test Screen.
•
MAIN MENU (F4)
to return to Main Menu 1.
Manometer: 0 mmHg Peak: 0 ZERO PR.> F7
PressDrop: 0 mmHg Time:00:00 RS PEAK.> F6
Leak Rate: 0 mmHg/min PRINT ..> F5
RESTART INFLATE SETUP MAIN MENU
F1 F2 F3 F4
6. Press
INFLATE (F2)
. The menu changes and
DEFLATE (F2)
becomes active
(see below).
Wait until the inflation stops. When
the test results are shown, press
DEFLATE
.
7. To reset the manometer and peak values to 0 mmHg, press
ZERO PR (F7)
and
RS PEAK (F6)
. Note that this also resets
the time to zero.
Manometer: 0 mmHg Peak: 0 ZERO PR.> F7
PressDrop: 0 mmHg Time:00:00 RS PEAK.> F6
Leak Rate: 0 mmHg/min PRINT ..> F5
RESTART INFLATE SETUP MAIN MENU
F1 F2 F3 F4
8. To repeat the test, press
RESTART (F1)
, which resets the results
to zero. Then, press
INFLATE
.
5-2
Summary of Contents for QA-1290
Page 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Page 6: ...This page intentionally left blank vi...
Page 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Page 30: ...This page intentionally left blank 4 8...
Page 36: ...This page intentionally left blank 5 6...
Page 51: ...Appendix A Tubing Kit Components A 1...
Page 52: ...This page intentionally left blank B 2...
Page 54: ...B 2...
Page 55: ...B 3...
Page 56: ...B 4...
Page 57: ...B 5...
Page 58: ...B 6...
Page 59: ...B 7...
Page 60: ...B 8...
Page 61: ...B 9...