5.2 Over-Pressure Cut-Off
Test
1. Advance to Main Menu Screen 2. From this screen press
Over-
Pressure Cut-Off Test (F7)
. This advances you to the test
screen.
Equipment Code : ......................> F7
Leak Test .............................> F6
BP Test ...............................> F5
MORE SET UP
F1 F2 F3 F4
Over-Pressure Cut-Off Test ............> F7
Manometer .............................> F6
Print header/footer ...................> F5
GO BACK SET UP
F1 F2 F3 F4
2. Press
SET UP (F3)
. This advances you to the Inflate Pressure
Screen.
(See below).
Manometer: 0 mmHg ZERO PRESSURE.> F7
Peak: 0 mmHg RESET PEAK ...> F6
PRINT ........> F5
INFLATE SET UP MAIN MENU
F1 F2 F3 F4
Inflate Pressure:XXX mmHg F7
F6
F5
STORE GO BACK MAIN MENU
F1 F2 F3 F4
3. The
Inflate Pressure (F7)
value must be entered. Otherwise the
test will fail. Use the alphanumeric keys to enter the pressure set-
ting to between 0 and 500 mmHg. Press the
Enter
key to save
the new setting(s). Use the
Delete
key to clear the current setting.
4. Press:
•
STORE (F1)
to store the settings in flash memory.
•
GO BACK (F2)
to return to the Over-Pressure Cut-Off Test
Screen.
5-3
Summary of Contents for QA-1290
Page 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Page 6: ...This page intentionally left blank vi...
Page 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Page 30: ...This page intentionally left blank 4 8...
Page 36: ...This page intentionally left blank 5 6...
Page 51: ...Appendix A Tubing Kit Components A 1...
Page 52: ...This page intentionally left blank B 2...
Page 54: ...B 2...
Page 55: ...B 3...
Page 56: ...B 4...
Page 57: ...B 5...
Page 58: ...B 6...
Page 59: ...B 7...
Page 60: ...B 8...
Page 61: ...B 9...