60/30 (40) HR:80 BPM PP:2.00 mmHg.> F7
80/48 (58) HR:80 BPM PP:2.00 mmHg.> F6
100/65 (77) HR:80 BPM PP:2.00 mmHg.> F5
MORE GO BACK START MAIN MENU
F1 F2 F3 F4
To select the desired preset use the
F5
,
F6
or
F7
key. The corre-
sponding line will flash intermittently to indicate the selected pre-
set. Below are the adult and neonate preset pressures.
Adult Preset Pressures
Diastolic//Systolic
(mmHg)
MAP (mmHg)
Heart Rate
(BPM)
Pulse Pressure
(mmHg)
60 / 30
40
80
3.00
80 / 48
58
80
3.00
100 / 65
77
80
3.00
120 / 80
95
80
3.00
150 / 95
114
80
3.00
200 / 140
167
80
3.00
255 / 195
215
80
3.00
Neonate Preset Pressures
Diastolic//Systolic
(mmHg)
MAP (mmHg)
Heart Rate
(BPM)
Pulse Pressure
(mmHg)
60 / 30
40
120
1.50
80 / 48
58
120
1.50
100 / 65
77
120
1.50
120 / 80
95
120
1.50
150 / 95
114
120
1.50
200 / 140
167
120
1.50
255 / 190
220
120
1.50
In the present pressure screen menu keys press:
•
MORE (F1)
to advance to the next preset screen.
•
GO BACK (F2)
to return to the Adult/Neonate Screen.
•
START (F3)
to advance to the BP Test Screen.
•
MAIN MENU (F4)
to return to Main Menu Screen 1.
4.22 Independent Test Parameters (F1)
. From the menu shown be-
low( from BP Test at Main Menu) select more and then
(F7)
to set
the desired blood pressure, heart rate or pulse amplitude and Ar-
rhythmias.
4-3
Summary of Contents for QA-1290
Page 1: ...QA 1290 User Service Manual 1 66 Firmware and higher QA 1290 NIBP Analyzer P N 17525 V 1 66...
Page 6: ...This page intentionally left blank vi...
Page 10: ...17520 User Manual QA 1290 P N 17525 Tubing Kit P N 17515 1 4...
Page 30: ...This page intentionally left blank 4 8...
Page 36: ...This page intentionally left blank 5 6...
Page 51: ...Appendix A Tubing Kit Components A 1...
Page 52: ...This page intentionally left blank B 2...
Page 54: ...B 2...
Page 55: ...B 3...
Page 56: ...B 4...
Page 57: ...B 5...
Page 58: ...B 6...
Page 59: ...B 7...
Page 60: ...B 8...
Page 61: ...B 9...