2.12 Method memory key <USER METH>
784 KFP Titrino
35
2.12 Method memory key <USER METH>
Management of the internal method memory with key
<USER METH>.
Select method name with keys <
←
> and <
→
>.
recall method:
Loads a method from the internal method memory into
the working memory.
store method:
Stores the method which is in the working memory in
the internal method memory.
delete method:
Deletes a method from the internal method memory.
The display texts of the Titrino are shown on the left
side. The values are the default values.
>recall method
method name:
Recall method
Recall method from the internal method memory to the
working memory (input of method name, which is
included in the memory).
If a method identification is entered which is not found
in the method memory, the selected value blinks.
>store method
method name:
Store method
Store method from the working memory to the internal
method memory (up to 8 ASCII characters).
If a method with an identical name is already stored,
you are requested if you wish to overwrite the old
method. With <ENTER> it is overwritten, with
<QUIT> you return to the entry.
USER METH
)
3
user methods
>recall method
>store method
>delete method