![Metrohm Titrino 784 KFP Instructions For Use Manual Download Page 36](http://html1.mh-extra.com/html/metrohm/titrino-784-kfp/titrino-784-kfp_instructions-for-use-manual_1779647036.webp)
2.10 Data output
784 KFP Titrino
33
Additional possibilities for report outputs
In addition to the reports which are printed at the end of the titration, various other reports
can be put out. There are 2 possibilities to select the reports:
1) <PRINT><
←
/
→
><ENTER>
Cursor is pressed repeatedly until the desired report
appears in the display.
2) <PRINT><keyX><ENTER>
key X is the key under which the appropriate data
are entered.
List of the "keys X":
Report
<Key X>
Configuration report
CONFIG
Parameter report
PARAM
Current sample data
SMPL DATA
Statistics report with the individual results
STATISTICS
All sample data from the silo memory
SILO
Operands C01...C19
C-FMLA
Contents of the <DEF> key
DEF
Contents of the method memory with details of the memory
requirements of the individual methods and the remaining bytes
USER METH
Complete report sequence of the last determination, as defined
under the <DEF> key in the method
REPORTS
Result display without printer
If you work without printer, we recommend to work with the standard character set for
result display (settings in key <CONFIG>, ">auxiliaries", see page 11). You will get the
complete information on your determinations: Calculated results, endpoints, messages
etc.
2.10.2 Display of the Curve
After the titration, the curve can be viewed.
Switch between "curve" and "result display" with key <CURVE>.
You can trace the curve with keys <
↑
> and <
↓
>. In
the text field on the left side of the curve the index of
the current measured value is displayed in the first line.
In the subsequent lines, the corresponding measured
values are shown.