2.4 Configuration, key <CONFIG>
784 KFP Titrino
9
2.4 Configuration, key <CONFIG>
Key <CONFIG> serves to enter device specific data.
The set values apply to all modes.
monitoring:
Monitoring of instrument validation, service interval and
printout of diagnostic report.
peripheral units:
Selection of printer, balance, stirrer control and COM
for manual report output.
auxiliaries:
e.g. setting of dialog language, date, time, type of re-
sult display.
RS232 settings COM1 and 2:
RS parameters for the COM's.
common variables:
Values of common variables.
The display texts of the Titrino are shown on the left
side. The values are the default values.
>monitoring
validation: OFF
time interval 365 d
time counter 0 d
service: OFF
next service YYYY-MM-DD
Monitoring functions
Monitoring the validation interval (ON, OFF)
Monitoring is carried out at the end of the titrations and
when the Titrino is switched on. If the monitoring
responds the message "validate instrument" appears.
The message vanishes with <CLEAR>. At the same
time the counter is reset to zero.
If "ON" has been set:
Time interval for validation (1...9999 d)
see also page 139.
Time counter (0...9999 d)
Counts the number of days since the last time the
counter was reset.
Monitoring the service interval (ON, OFF)
Monitoring is carried out after the Titrino has been
switched on. If the monitoring responds the message
"Service is due" appears. The message vanishes with
<CLEAR>.
If "ON" has been set:
Date of next service (YYYY-MM-DD)
configuration
>monitoring
>peripheral units
>auxiliaries
>RS232 settings COM1
>RS232 settings COM2
>common variables
CONFIG