![Metrohm Titrino 784 KFP Instructions For Use Manual Download Page 25](http://html1.mh-extra.com/html/metrohm/titrino-784-kfp/titrino-784-kfp_instructions-for-use-manual_1779647025.webp)
2.6 Parameters, key <PARAM>
784 KFP Titrino
22
Mechanisms of the KFT parameters in Ipol mode
•
The position and curve characteristics of the line between the ranges of excess water
or excess iodine depends on the type of sample and the ingredients of the working
media.
•
The endpoint has to be set close to the range margin, but always within the iodine
excess range. If the endpoint is set too close to the limit, an overdose of KF reagent
may be dispensed. The steeper the curve at the preset polarization current I(pol), the
more delicate it is to titrate to a steady and reproducible endpoint. For the titration of
troublesome samples, the particular conditions must be optimized by trial and error.
Always mind the color of the working medium at the end point as your guideline.
Remark:
In most cases the preset default parameters provide precise and
reproducible results.
•
If you select negative values for a parameter, you should choose negative values for all
other parameters too in order to have compatible parameter sets.
•
In KFT Upol mode, all parameters work in a similar manner as in KFT Ipol mode.
U/mV
I/
µ
A
+
+
I(pol)
EP at U
Titration
(Back titration)
Titration
(direct titration)
Excess H
2
O
Excess Iodine-
darker
solution