2.7 Result calculations
784 KFP Titrino
26
RS1 text RS1
RS1 decimal places 2
RS1 unit: %
RS1 limit control: OFF
RS1 low lim. 0.0
RS1 up lim. 0.0
RS1 L13 output: OFF
Text for result output (up to 8 characters)
Text input see page 8.
Number of decimal places for result (0...5)
Selection of result unit (%, ppm, g/L, mg/mL, mol/L,
mmol/L, g, mg, mL, mg/pc, s, mL/min, no unit or up to
6 characters).
Limit control for the result (ON, OFF)
The limits are checked each time a result is calculated.
If "ON" has been set:
Lower limit (0.0...999 999)
Upper limit (0.0...999 999)
Sets line L13 of the remote socket (OFF, active, pulse)
if the result lies outside the limits.
Enter next formula, e.g. for RS2.
Meaning of the calculation variables CXX:
C00
Sample size, see page 37.
C01...C19
Method specific operands, see page 27. They are stored with the
method in the method memory.
C21...C23
Sample specific operands, see page 37ff.
C26, 27
Mean values from silo calculations, see page 41ff.
C30...C39
Common variables.
C40
Initial measured value of the sample.
C41
End volume.
C42
Determination time.
C43
Volume drift for KFT with conditioning.
C44
Temperature.
C45
Dispensed start volume.