2.6 Parameters, key <PARAM>
784 KFP Titrino
20
Titration sequence of KFT
After the start, the activate pulse is outputted and the stir-
rer switched on.
The start delay time is waited off.
If conditioning is on, the sample solution is titrated until
the EP is reached. The display shows then
drift OK 2.3
µ
l/min
or
KFT conditioning
The vessel is now conditioned. The titration can be started
with <START>.
The sample identifications and the sample size are re-
quested.
Pause 1 is waited off, the start volume is dispensed and
pause 2 waited off.
The titration is executed. If the extraction time is not yet
over when the endpoint has been reached, the titration will
only be terminated after the extraction time is over.
The stirrer is switched off when there is no conditioning.
Calculations are carried out.
Data are outputted.
Conditioning is carried out.
(Activate pulse)
(Stirrer ON)
<START>
(Start delay)
(Start conditions)
(Request ident.)
(Request smpl size)
(Extraction time)
Titration with test of
stop criterion
Calculations
Data output
(Reconditioning)
(Preconditioning)
(<START>
(Activate pulse)
(Start delay)
(Stirrer OFF)