4.2 Remote control commands
756/831 KF Coulometer, Instructions for Use
86
I
nfo
.A
ctualInfo
.O
utputs
.Cl
ear $G
Status sends the current status of the I/O lines, Change sends the information
regarding whether a change in status of a line has taken place since the last
clearing, Clear clears the change information. For the output, there is a con-
version from binary to decimal, e.g.
| 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 0 | 1 | 0 | 1 | 0 |
Line No.
| 13| 12| 11| 10| 9 | 8 | 7 | 6 | 5 | 4 | 3 | 2 | 1 | 0 |
Output: 2
1
+ 2
3
= "10"
1 means ON or change; 0 means OFF or no change.
The lines are assigned as follows (see also page 131):
Inputs
:
Outputs
:
0
Start (pin 21)
0
Ready (pin 5)
1
Stop (pin 9)
1
Cond. ok (pin 18)
2
Enter (pin 22)
2
Titration (pin 4)
3
pin 10
3
EOD (pin 17)
4
pin 23
4
not used (pin 3)
5
pin 11
5
Error (pin 16)
6
pin 24
6
Activate, line L6 (pin 1)
7
pin 12
7
Pulse for recorder (pin 2)
8
Connected remote box (pin 6)
9
not used (pin 7)
10
not used (pin 8)
11
Change reagent (pin 13)
12
Smpl size out (pin 19)
13
Result out (pin 20)
4.2.2.70. I
nfo
.A
ctualInfo
.A
ssembly
.C
yclNo read
only
I
nfo
.A
ctualInfo
.A
ssembly
.I
read
only
I
nfo
.A
ctualInfo
.A
ssembly
.M
eas read
only
I
nfo
.A
ctualInfo
.A
ssembly
.P
ot read
only
I
nfo
.A
ctualInfo
.A
ssembly
.IP
ulse read
only
I
nfo
.A
ctualInfo
.A
ssembly
.B
ur
.V
read
only
I
nfo
.A
ctualInfo
.A
ssembly
.B
ur
.C
lear $G
$Q sends the current values.
.CyclNo: Cycle number of the voltage measurement cycle, e.g. "127". From
the cycle number and the cycle time (see 4.2.2.77), a time frame
can be set up. The cycle number is set to 0 on switching on the
instrument and on every start. It is incremented as long as the in-
strument remains switched on.
.I:
Total charge in mA·s, e.g. "667.48".
.Meas:
Measured value in mV at the indicator electrode, e.g. "104.2".
.Pot:
Voltage at generator electrode.
0 means "undefined", 1 means <14 V, 2 means 14...28 V, 3
means >28 V.
.IPulse:
Current of actual pulse.
1 means 100 mA, 2 means 200 mA, 3 means 400 mA.
.Bur.V:
Dosed volume of connected Dosino in ml, e.g. "5.234".
.Bur.Clear: $G clears the volume counter.
4.2.2.71. I
nfo
.A
ctualInfo
.T
itrator
.C
yclNo read
only
I
nfo
.A
ctualInfo
.T
itrator
.W
ater read
only