7.4 User methods
756/831 KF Coulometer, Instructions of Use
139
Blank value method, parameters:
'pa
756 KF Coulometer
5.756.0010
date 1997-11-19 time 17:56 0
BLANK
774-Blk
parameters
>control parameters
EP at U
50 mV
dynamics
70 mV
max.rate
max. ug/min
min.rate
15 ug/min
stop crit:
rel.drift
rel.drift
5 ug/min
>titration parameters
pause
0 s
extr.time
180 s
start drift
10 ug/min
I(pol):
10 uA
electrode test:
ON
temperature
25.0 °C
time interval
2 s
max.titr.time
OFF s
>statistics
status:
ON
mean
n= 3
res.tab:
original
>preselections
drift corr:
auto
req.ident:
OFF
req.smpl size:
OFF
smpl unit:
g
limit smpl size:
OFF
text id1
id1 or C21
text id2
id2 or C22
text id3
id3 or C23
cell:
no diaph.
generator I:
400 mA
oven:
COM1
activate pulse:
OFF
------------