6.4 Connecting the 774 Oven Sample Processor
756/831 KF Coulometer, Instructions for Use
120
6.4.1 Equipping the titration vessel with the Oven Sample Proc-
essor
Lead the tip of 6.1830.010 heatable outlet tube into the titration vessel with the aid of
6.1446.170 stopper.
Outlet heater
If in addition a Dosino is to be used for aspiration then the aspiration tip is inserted into
the joint opening and the gas inlet is fitted with A.254.0102 seal which is placed in the
screw cap instead of the septum; the screw cap is then screwed down; see illustration on
page 118.
If you require an additional opening for injections then 6.1465.320 titration vessel is avail-
able; it has two side-mounted screw threaded openings.
Stopper with
6.1446.170 O-ring
Generator
electrode
with drying
Connection for
outlet heating
Transfer hose with
outlet heating
6.1830.010
Indicator
electrode
Septum stopper
6.1464.320
Titration vessel
728 Mag-
netic stir-
Fastening screws