7010 TOC ANALYZER USER MANUAL
26
2.5.1 COPPER FILTER
The copper filter
(02)
is located in the electrical enclosure immediately before the IR detector inlet. It is a
plastic container filled with copper wool. The gas leaving the dryer tubing is forced to go through this device to
prevent corrosive effects due to gases like chlorine or chlorine dioxide that could be generated in the oxidation
stage.
The sodalime filter
(05)
is located on the left side of electrical enclosure. It’s a plastic container full of sodalime
and it absorbs the carbon dioxide from atmospheric air providing the analyzer with CO
2
free air for its
processes.
2.5.2 SODALIME FILTER