7010 TOC ANALYZER USER MANUAL
24
2.4.6 DRYER
The dryer
(05)
is positioned between the BYPASS VALVE and the IR detector. It consists of a coil of two
concentric tubes. In the internal tubing flows the gas stream to dry. This tubing is water vapour permeable so
that the humidity passes to the external tubing. In the external tubing there is a countercurrent purge gas flow
that removes the water vapour. The dryer prevents water condensation inside the IR cell
CAUTION : do not overtighten or twist gas dryer ends when installing or servicing, as this will restrict
the airflow.
The Bypass valve shown in the diagram as
(03)
is activated during conditioning and cleaning phases.
It vents the gas stream avoiding it to flow through the IR cell when not strictly necessary.
2.4.7 BYPASS VALVE