5.2 COPPER WOOL REPLACEMENT
The halogens filter is located into the electrical enclosure ( please read the hazards and dangers list
in section 1). This operation should be made by qualified personnel who have been fully trained and
has professional experience to avoid electricity hazards and dangers.
A)
put the analyser in standby and open the analyzer electrical enclosure
B)
disconnect the inlet and outlet tubing of the filter from their fittings
C)
remove the filter plastic body from its support clamp
D)
unscrew top and bottom caps from filter body
E)
with extreme care, using a proper tool, pull out the used wool from ther plastic cylinder
F)
replace the used wool with new wool and press it into place to form a compact pod.
G)
screw back on the top and bottom caps
H)
put the filter in the clamp, connect fittings and turn on line the analyzer