7010 TOC ANALYZER USER MANUAL
23
2.4.4 UTUBE
The UTube
(13)
is located in the middle position of liquids enclosure. It is a U shaped glass device, with two
inlet and two outlet points.
It separates the analyzed liquid part of the sample coming from UV reactor from the gaseous stream directed
to the infrared analyser. It also drains the exausted sample and vents the sparging gas coming from the top of
the scrubber.
The gas mixture coming from the oxidation stage is driven by the carrier gas to the drying devices through the
upper right outlet of the gasliquid separator, towards to the glass condenser
(02)
.
The Glass condenser
(02)
is positioned between U tube and the bypass valve. The glass condenser uses the
temperature difference between its glass body, cooled by a fan, and the hot treated sample coming from UV
reactor.
2.4.5 GLASS CONDENSER