Chapter 7
Diagnosis
The FTC300 has several integrated diagnosis and test functions that can be accessed through the
diagnosis-menu. The menu provides the following functions:
• A parameter menu in which device-internal parameters/variables can be read out
• An error menu in which pending errors are listed
For devices operated in “safety mode“, the operator code is needed to access the diagnosis menu.
By default the operator code is set to 111.000.
7.1
Parameter Menu
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Parameter
Errors
Compound Raw
set: 3000.00
------------
P001 OK
Figure 7.1: parameter menu for advanced diagnosis
The configuration of the FTC300 is defined by a list of internal parameters. The “Parameter“ menu
gives read-only access to these parameters. This may help an experienced user to diagnose mal-
functions caused by wrong settings. The parameter menu allows you to scroll through the entire
parameter list. Contact Messkonzept for detailed information on the listed parameters. The first dis-
play line contains the name of the parameter, in the second line the parameter value is shown. The
last line shows the parameter index. To move forward in the list, press <ENTER>, to move back-
wards, press <Up>. To leave the parameter menu mark <OK> by pressing <RIGHT> and confirm
with <ENTER>. Some parameters can be changed in the expert setup, see Section 8.5 for more
information.