Page 31 of 58
8.2.3
Response Time Setup
A1 A2 Er
0.30
[%]
H2/N2
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Operation
Calibration
Diagnosis
Setup
Unit -> Vol%
H2 in N2
T90 Response
List for MGM
T90 Response
set: 1.00000
new: 1.00000
[s]
ESC
/OK
Figure 8.4: T90 response time setup
In the menu “T90 Response“ the response time of the exponential filter can be adjusted. The filter
reduces the influence of fast variations of the raw signal (signal noise smoothing). The numerical
value for the T90 response time is given in the unit of seconds. A range between 0s and 100s is
permitted for this value, reasonable values lie between 0.5s to 10s. You can change the numerical
value by navigating through the digit position using the <RIGHT> key and changing it using the <UP>
key. Confirm by pressing the <UP> key until ”OK” is selected. The changed value can be discarded
by selecting ”ESC” instead. Setting “T90 Response“ to 0.0 turns the exponential filter off.
The true response time is influenced by the gas exchange time which depends on the
pneumatic installation and the flow rate of measuring gas. A gas flow of 80l/h leads to a
gas exchange time of under 0.5s measured from the gas inlet of the device.
Note that the response time in your process depends greatly on the upstream volume
before the FTC300. Shorter and/or thinner tubing will benefit response times.
The T90-time is the time, in which a sudden change of the measurand (e.g. the gas
concentration) reaches 90% of it’s final value.
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf