8.3
Relay Setup
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
Figure 8.6: selection of configurable relays
The menu “Relay Setup“ gives access to the configuration of the two process alarm relays “Relay 1“,
“Relay 2“ and the status alarm relay “Common Relay“. The thresholds and hysteresis for “Relay 1“
and “Relay 2“ can be set. The status relay indicates a malfunction of the device, but can also indicate
the threshold alarms. Each relay may be configured as failsafe or not failsafe.
The threshold relays are indicated by “A1“ respectively “A2“ in the first display line of the
operation screen. The common relay is indicated by “Er“ in the first display line of the
operation screen and also by the flashing red LED on the front panel.
8.3.1
Relay 1 Mode
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Relay 1
Relay 2
------------
Common Relay
R1 Mode ***
Failsafe
Cal. active
ESC/OK
R1 Mode L/H
R1 Threshold
R1 Hyst.
ESC/OK
R1 Mode OFF
------------
------------
ESC/OK
R1 Mode H/L
R1 Threshold
R1 Hyst.
ESC/OK
Figure 8.7: relay mode setup
Relay 1 can be operated in different modes described below. Selecting “Relay 1“ and pressing
<ENTER> gives access to the different modes: