Page 42 of 58
8.4.3
Analog Output 2
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Current Loop
Analog Out 1
Analog Out 2
Adjust Loop
AOut2-Param
AOut2-Offset
AOut2-Gain
AOut2-Param
set: 0.00000
new: 0.00000
ESC
/OK
AOut2-Param
AOut2-Offset
AOut2-Gain
AOut2-Param
AOut2-Offset
AOut2-Gain
AOut2-Offset
set: 0.00000
new: 0.00000
ESC
/OK
AOut2-Gain
set: 0.00000
new: 0.00000
ESC
/OK
Figure 8.17: Setup of Analog Output 2
All “Analog Out 2“ settings can be made analogously as for “Analog Out 1“ described before.
8.4.4
Current loop calibration
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Current Loop
Analog Out 1
Analog Out 2
Adjust Loop
Reading...
set: 19.5000
new: 19.5000
[mA]
ESC
/OK
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Figure 8.18: Calibration of the current loop
In the menu “Adjust Loop“ the current loop can be calibrated. Opening the menu “Adjust Loop“ sets
the current loop output to 19.500mA. Measure the resulting real current at the output (e.g. with
a multimeter) and enter the measured value in the third display line, then confirm with “OK“ and
<ENTER>. After successful adjustment the multimeter should show a value of 19.5mV.