Page 12 of 58
“Basic range“ is the largest possible measuring range and is set by default.
The
linearization is performed over the basic range. The smallest measuring ranges at the
beginning and the end of the basic range are facilitated through specific calibration. The
smallest possible range between the basic range and the smallest ranges at the end
beginning and the end of the range can be estimated by linear Interpolation.
Measuring ranges can be set up in the “Current Loop“ menu, see Section 8.4.1
The Multi Gas Mode (MGM) is a configuration that allows for the consecutive
measurement of different gas pairs. The gas pair can be switched through the control
panel or via the RS232-interface. Gas pairs labeled “Yes“ in Table 2.1 are commonly
used. Gas mixtures labelled “On request“ can also be implemented upon request.