Page 14 of 58
6mm
52 mm
20 mm
26 mm
80 mm
65 mm
144 mm
132 mm
4.5mm
Figure 3.2: Mounting holes shown from the reverse side of the housing
If you are planning to lead flammable or toxic gases into the device, the device must be
installed in a well ventilated area. All devices undergo a leakage test during production,
nevertheless a limited release of small gas quantities is possible.
3.3
Gas Ports
On the bottom of the FTC300 housing two tubes with 6mm outer diameter for gas connection are
located. They are labeled with “GAS IN “ and “GAS OUT“.
For low requirements regarding gas tightness and resistance to pressure the tubes can be used as
hose connector. For permanent gas and pressure tightness compression fittings are recommended
(e.g. by “Swagelok“©).
After connecting the device a leakage test should be performed (especially when working with
flammable and/or toxic gases).