![MessKonzept FTC300 Operating Manual Download Page 44](http://html.mh-extra.com/html/messkonzept/ftc300/ftc300_operating-manual_1775512044.webp)
Page 43 of 58
8.5
Expert Setup
A1 A2 Er
0.30
[%]
H2/N2
Operation
Calibration
Diagnosis
Setup
Instr. Set.
Relay Set.
Output Set.
Expert Set.
Expert Code
set: 0.00000
new: *******
ESC/
OK
Figure 8.19: Expert Setup menu
The Expert Setup provides a couple of functionalities that should only be used by an advanced user
or expert:
• set parameters
• reset to factory settings
• change the “Operator Code“ and the “Expert Code“
• swap between “Normal Mode“ and “Safety Mode“
• simulate alarms and analog outputs
The settings explained here are for advanced users or experts and should not be
entered by normal operators. It is in the responsibility of experts to set the parameters
properly. The default expert code is “222.000“.
8.5.1
Parameter
Parameter
Access Modes
Reset Funct.
I/O Test
LCD Contrast
set: 25.0000
Change ?
P009 OK
LCD Contrast
set: 25.0000
Change ?
P009 OK
LCD Contrast
set: 25.0000
new: 25.0000
ESC
/OK
Compound ppm
set: 2105.13
------------
P009 OK
Figure 8.20: Parameter change in expert mode
The configuration of the FTC300 is represented by an internal list of parameters. These param-
eters govern all settings and functions of the device. In the expert-menu’s parameter list press
<ENTER> to scroll forward through the list and backwards by pressing <UP>. Some parameters
cannot be changed (e.g. sensor information such as “Compound ppm“, see Figure above) others
can be changed (e.g. “LED Contrast“, see Figure above). All changeable parameters are indicated
by “Change?“ in the third line of the display. Selecting and clicking “Change?“ opens a submenu in
File name:
FTC300 Manual 1.09KD140704CWI1V05.pdf