nadac.merging.com
Page 73
11.
Wait until the webpage says “
Update Successful!
” (Figure 96) and the unit’s front panel display shows
Please reboot
. The power button light will also change to a steady green.
Shutdown the unit by clicking on the
Reboot Device
button in the firmware update screen - Figure 96. The
unit will restart automatically afterwards.
12.
You can verify the firmware version installed by going into the unit’s Setup menu and selecting Info – see
page 37. The firmware version will be shown at the bottom of the display.
If you had to change your computer IP address in order to perform the firmware update, you may now re-
set it again to your preferred IP address using the units’ Network menu – see page 35.
If the Firmware update procedure fails, the screen shown in Figure 97 will usually be displayed.
Press the
Try Again
button to repeat the update process and follow these instructions starting at
step 7.
Alternatively use your browser’s go back button to return to the screen shown in Figure 92, the
firmware again and re-update it. In this case re- start again at step 7.
Avoid powering down the unit if the firmware update has failed.
If you do not follow this procedure the unit may no longer start up and it may be necessary to
return to a local service centre or the factory.
If you cannot update the firmware, please contact your local Merging Technologies reseller for assistance.
Figure 96. Firmware update successful.
Figure 97. Firmware update
Try Again
button.
Summary of Contents for MERGING+NADAC
Page 1: ...REFERENCE GUIDE ...
Page 75: ...nadac merging com Page 74 ...