Connection to supply of untreated water (normal city water) 

Install the MELAdem

®

45 reverse-osmosis system in accordance with the installation instructions provided 

with the MELAdem

®

45 system.   

Connect the MELAdem

®

45 reverse-osmosis system to a supply of untreated water (with a 6 

 1 hose) to a 

tap from the normal tap water supply (Item 10), by means of a 3/4" inside thread / hose connection (hose 
adapter 6 

 1, see Item 16), as provided from the Drain Set. If a separate water tap is not available for con-

nection of the water-treatment system to an existing washup sink, then the existing angle-type non-return 
valve may be replaced by another valve with 3/4" outside-thread connection.   

Installation of the leak monitor (optional) 

If the water-treatment system is connected to a 3/4" normal water tap, then a leak monitor (shuts off running 
water in case of leaks) can be installed as option.  The shutoff valve (Item 9) must then be directly connected 
to the normal city water tap by means of the union nut.  The hose adapter (Item 16) must be screwed on at 
the outlet of the solenoid valve, and the untreated tap water connection hose must be connected to the ME-
LAdem

®

45 reverse-osmosis system.  Plug the leak monitor (shuts off running water) (Item 25) into an elec-

trical socket, and install the electrical leads from the shutoff valve and the leak sensor (Item 27). 

Connection of the control unit for external water refilling (external solenoid valve) / pure-water con-
nection 

Screw the control unit for external water refilling (external solenoid valve, Item 2) onto the connection for 
water feed [39] on the rear side of the autoclave.  Connect the power cable of the valve to the socket of the 
solenoid valve for water feed [76] on the autoclave.  Connect the incoming-feed hose from the MELAdem

®

45 

reverse-osmosis system to the threaded fitting of the control unit for external water refilling, at the external 
solenoid valve, Item 2.   

Wastewater connection 

Use the union nut to connect the Y-fitting of the Drain Set (Item 11) to the existing 1" branch of the washup-
sink trap.  

Important:

  Install the two back-pressure valves, with the aid of the guide grooves provided, into 

the free ends of the Y-fitting. 

Connection of the autoclave overflow 

Use the elbow coupling to connect the overflow hose (Item 23) to the overflow fitting [68] of the autoclave, 
and to the hose nozzle connector and union nut (Item 14) on the Y-fitting.  

Important:

  When installing the 

overflow hose, make sure that the hose passes 

continuously downward

, without any upward curve. 

Wastewater connection of the MELA

dem

®

45

 reverse-osmosis system 

Connect the wastewater hose of the MELAdem

®

45 reverse-osmosis system to the free branch of the drain 

Y-fitting, by means of the hose adapter for the 6 

 1 hose (turning piece with G1/4", inside threads, see Item 

12), by using the screwed-in hose threads, 6 

 1 (Item 17), and the union nut. 

Connection of overflow for the condensate container (optional) 

The user can install an optional overflow at the condensate container, to provide automatic emptying of the 
condensate container.  Remove the dummy plug from the overflow fitting of the condensate container, and 
connect the overflow tube (item 18) with its pipe elbow to the overflow fitting.   

Use the Y-fitting (Item 13) and the threaded hose connection for hose, 8 

 1 (Item 15), of the Drain Set to 

connect the overflow tube to the drain, together with the drain connection of the MELAdem

®

45 reverse-

osmosis system.   

Install the overflow tube in such a manner that the tube passes 

continuously downward

, without any up-

ward curve.  The maximum permissible level difference between the drain fitting on the trap and the overflow 
fitting on the condensate container is + 0.5 meters. 

Summary of Contents for Vacuklav 31

Page 1: ...ucceeded in becoming a leading manufacturer of sterili zation equipment More than 300 000 MELAG units sold throughout the world testify to the exceptional quality of our sterilizers As all other MELAG...

Page 2: ...nitoring 10 2 4 2 Periodic testing every six months 10 2 4 3 Maintenance service of the autoclave 10 3 Installation and initial startup 11 3 1 Setting up the autoclave 11 3 1 1 Transport ribbons 11 3...

Page 3: ...ransport placing the autoclave back into operation 26 6 Malfunctions 27 6 1 Malfunctions without error display 27 6 1 1 Display of the operational state 27 6 1 2 A program does not start 27 6 2 Malfun...

Page 4: ...r gasket 3 Display 11 Twist door handle 4 Front panel with front film scale 12 Door locking nut 5 Enclosure cover above the water storage tank 13 Door 6 Door locking spindle 14 Door contact 7 Housing...

Page 5: ...error 36 Arrow pointing to power switch 1 1 3 View of rear _________________________________________________________________________ 38 Sterile filter 74 Data interface for printer 39 Connection for w...

Page 6: ...into the program sequence It ensures excellent drying results within a short drying period This optional supplementary drying enables the operator to extend the standard drying time in order to achie...

Page 7: ...ruments to be sterilized are thoroughly cleaned before sterilization in order to prevent contamination from being separated from the instruments during sterilization and from clog ging the filters and...

Page 8: ...storage tank for example by using a small brush with hot water and a grease cutting additive Or use a non raveling cloth soaked in alcohol Then thoroughly rinse out the water storage tank with water a...

Page 9: ...iO2 0 1 mg kg 1 mg l Iron 0 1 mg kg 0 2 mg l Cadmium 0 005 mg kg 0 005 mg l Lead 0 05 mg kg 0 05 mg l Rest of heavy metals ex cept iron cadium and lead 0 1 mg kg 0 1 mg l Chlorides Cl 0 1 mg kg 2 mg l...

Page 10: ...m the results of sterilization on an official test log Spore tests to confirm the effectiveness of sterilization of an autoclave provide valid results only in conjunc tion with the sterilization progr...

Page 11: ...the autoclave is to use a spirit level carpenter s level to cor rectly align the unit The spirit level must first be held up flush against the flange of the chamber to ensure that the entire unit is v...

Page 12: ...onnected to the condensate container 80 43 One way drain system 73 Connection for drain to con densate container 80 Condensate container 81 Connection for drain from autoclave 82 One way drain system...

Page 13: ...rts 23 Then start the de sired program again 3 3 2 Option connection from condensate container to a building drain___________________________ An option is automatic draining of the condensate containe...

Page 14: ...the water storage tank over the maximum filling mark MAX 3 5 2 Automatic water refill_________________________________________________________________ As delivered by the manufacturer the Vacuklav 31...

Page 15: ...water as supply into the MELAdem 45 reverse osmosis system drain valve with pipe vent non return valve 1 2 x 3 4 Drain water connection fitting for drain water outflow from the MELAdem 45 reverse osmo...

Page 16: ...ng feed hose from the MELAdem 45 reverse osmosis system to the threaded fitting of the control unit for external water refilling at the external solenoid valve Item 2 Wastewater connection Use the uni...

Page 17: ...eaks Test for overflow of the autoclave This test is carried out by blocking the float switch in the water storage tank of the autoclave The user must then check to ensure that the inflowing water is...

Page 18: ...ece in 01045 For more detailed information on the connection of water treatment systems by other equipment manufac turers please consult your specialist MELAG dealer and mention the key words External...

Page 19: ...time the autoclave system will balance itself The test itself lasts ten 10 minutes If the system determines a leakage rate of 1 3 mbar min the test is considered successful and the follow ing completi...

Page 20: ...n Program select 18 When you press start stop 19 you have installed the printer 3 9 VDE stipulations ________________________________________________________ According to VDE stipulations presently va...

Page 21: ...amage to the autoclave and to the instruments being sterilized 4 4 Loading the autoclave ___________________________________________________ The operator can use either trays containers or standard tr...

Page 22: ...lave chamber The display will show the progress of the fractional evacuation as follows Display line 1 Sub atm pulsing Display line 2 0 2 bar 60 C Fractional evacuation will be repeated several times...

Page 23: ...C Display line 1 Immediat removal Display line 2 Press STOP 4 13 Supplementary drying __________________________________________________ The Vacuklav 31 has been provided with the supplementary drying...

Page 24: ...ing are various print and data saving possibilities which can be initiated by pressing the button for printing 22 and the and buttons 17 Pressing the button for printing once Display line 1 Immed prin...

Page 25: ...release The display then notifies the user of the type of mal function In addition if a printer is connected to the autoclave it will print out the malfunction report and will also show the values fo...

Page 26: ...t and allow the autoclave to cool down Remove the hoses and tubes from the rear of the autoclave Open the tap for draining the water storage tank and allow the water to run out of the tank If you wish...

Page 27: ...cket carrying the proper line voltage If the plug is in the socket and if the power switch is on the autoclave display will light up If line voltage is provided and if the power switch is on but if th...

Page 28: ...ralized water aqua dem aqua dest Press the button for acknowledg ing error reports 23 Then start the program again MALFUNCTION Temp too high The malfunc tion signal lamp lights up 35 The steam tempera...

Page 29: ...cartridge module The malfunction signal lamp lights up 35 The conductivity of the water is too high You can still start your program however by pressing the start stop button 19 Empty the water stora...

Page 30: ...ed instruments 19 24 15 min 10 min Textiles 33 41 15 min 10 min Disinfection 20 27 15 min 10 min Bowie Dick Test approx 26 min 5 min 5 min Vacuum test approx 16 min 7 1 3 Electrical connections consum...

Page 31: ...ys Max of 1 5 kg of textiles Possibilities of loading Tray racks with a max of five 5 MELAG trays Tray racks with a maximum of three 3 standard tray cassettes or three 3 MELAG trays MELAG sterilizatio...

Page 32: ...heating these objects and the containers As a result of this process the steam undergoes a change of state from gaseous to liquid This liquid conse quently appears as condensate on the instruments and...

Page 33: ...acked vertically fig B This vertical configuration ensures that channels can form between the textile folds for the air to flow out and for the steam to flow in Do not stack textiles horizontally fig...

Page 34: ...agents not only protect the instruments themselves they also protect the microorganisms on the instruments 8 4 Loading the autoclave ___________________________________________________ Do not sterili...

Page 35: ...ely to the bottom of the chamber The best solution arrange several sterilization containers of the same size one on top of each other so that the condensate can run down the side walls of the steriliz...

Page 36: ...r evaporate the rest of the condensate after the sterilization process has been completed German Standard DIN 58953 Part 7 Section 7 contains the following information concerning moisture which may re...

Reviews: