6-Using the Multimeter
36
6.6 Taking a Relative Measurement
A currently measured result relative to the defined reference value is displayed in a relative
measurement.
The following example shows how to take a relative measurement. At first, it is required to
acquire a reference value.
1.
Press
R
key and
R
is displayed on the top side of the screen.
2.
Press the
AUTO SET
key till
C
appears at the top of the screen.
3.
Plug capacitance extended module in capacitance measure jack.
4.
When the reading leveling off, press
F2
key and
△
is displayed on the top side of the screen.
The saved reference value is displayed below
△
.
5.
Plug capacitance, the displayed major reading on the screen is actual the capacitance value.
Refer to figure 32
figure 32: Relative Measurement
6.7 Selecting Automatic/Manual Range Adjustment
The defaulted range mode of the instrument is automatic range. To switch to the manual range,
perform the following steps:
1.
Press
F1
key and
MANUAL
is displayed on the top left side of the screen to enter the
manual range mode.
2.
Under the manual range mode, the measuring range is increased by a stage when pressing
F1
key each time, and when reaching the highest stage, it jumps to the lowest stage by pressing
F1 key once again. To multiply the reading of dial pointer by multiplying power and the unit
of main reading on the screen will get the measurement result.