3 iNSTRUMENT DESCRiPTiON AND ACCESSORiES
3
I
nstrument description and
Accessories
3.1 Top panel
1
2
4
5
11
13
12
7
9
8
10
6
3
1.
TXL STOP
Output used for stop discharging from an external
device (TXL). Galvanically isolated.
2.
SERVICE
Connector for service purposes only.
3.
ALARM
Output equipped with a relay contact for trigger-
ing an external alarm device.
4.
DC OUT
9 V output for external current clamp.
5.
I EXT
≤
1 V
input used to measure current in an external path
by means of a DC clamp-on current probe or a
current shunt.
6.
Display
Touch screen 7"
7.
BVM1, BVM2
USB connections for BVM units.
8.
USB connection
For USB memory stick.
The USB memory stick must be FAT32 formatted.
9.
Ethernet connection
For service of the instrument.
10.
EMERGENCY STOP
Push to stop.
The connection to the test object (battery bank) is
interrupted in the same way as if mains would fail.
in addition, a separate signal will stop the electron-
ics and the discharge process stops.
The emergency stop works even if the electronics
fail.
Reset the button by turning it right
11.
Control knob
For entering settings etc. Press to confirm a setting.
12.
Buzzer
For alarms.
13.
ON/OFF switch
Summary of Contents for TORKEL 910
Page 2: ......
Page 13: ...CS033663DE ZP CS01E TORKEL 900 series 13 4 Menu system...
Page 27: ...CS033663DE ZP CS01E TORKEL 900 series 27 6 External current measurement...
Page 31: ...CS033663DE ZP CS01E TORKEL 900 series 31 7 Alarm function...
Page 33: ...CS033663DE ZP CS01E TORKEL 900 series 33 8 How to obtain the desired current...
Page 41: ...CS033663DE ZP CS01E TORKEL 900 series 41 10 Troubleshooting...
Page 43: ...CS033663DE ZP CS01E TORKEL 900 series 43 11 Calibration...
Page 45: ...CS033663DE ZP CS01E TORKEL 900 series 45 12 Specifications...
Page 49: ...CS033663DE ZP CS01E TORKEL 900 series 49 12 Specifications...
Page 51: ......