CS033663DE
ZP-CS01E
TORKEL 900-SERiES
37
9 OPTiONAL EQUiPMENT
1]
Set the range selector switch to the desired
position.
2]
Connect as shown below.
3]
Connect the control wires between the
TXL
CTRL
output on TORKEL and the
CONTROL
IN
input on the TXL.
If two or more TXLs are to be used, provide
a connection between the
CONTROL OUT
output on the first TXL and the
CONTROL IN
input on the second TXL, etc.
4]
Connect the TXL to the mains voltage
5]
Switch on the TXL.
Testing
WARNING
See the chapter "2.2 Safety instructions"
on page 8. for safety precautions.
1]
Proceed in the same way as set forth in the
chapter headed "5.1 Preparations for test-
ing" on page 20 but before you start
TORKEL you must set switch F1 to upper (ON)
position on the TXL. (You must do this manu-
ally.)
2]
Start TORKEL by pressing
3]
When TORKEL is stopped, manually or by any
stop condition, the TXL will also shut off the
discharge current.
9.2 BVM - Battery Voltage
Monitor
The BVM (Battery Voltage Monitor) is a battery volt-
age measurement device that is used for monitoring
of cell voltages and battery blocks in battery banks
commonly found in electrical power sub-stations, tel-
ecom facilities and computer data center UPS systems.
in conjunction the TORKEL unit, and test data man-
agement software, such as PowerDB or TORKEL Win,
the BVM enables a completely automated battery
bank capacity test to be performed.
BVM is available in three kits:
▪
BVM150 with 16 BVM units
▪
BVM300 with 31 BVM units
▪
BVM600 with 61 BVM units
1
3
2
1.
Control cable – Output socket
2.
Control cable – input socket
3.
Battery sense – Dolphin clip
The BVM is designed in modular form where one BVM
device is used for each battery or “jar” in the string
to be tested. it measures the voltage for each battery
cell and each unit can handle up to 20 V. The BVMs
connects to the batteries in a “daisy-chain” fashion,
thereby providing easy and economical expandability
to meet the testing requirements for small-to-large
battery bank systems. Up to 240 BVMs can be daisy-
chained in a single battery bank under test.
The included dolphin-style battery clamp can be easily
removed and exchanged with different styles of stand-
ard banana plug clamps and/or extension cables to
accommodate any battery connection requirement.
There is an Activity LED on the BVM that flashes when-
ever an individual BVM transmits data back to the data
acquisition host computer. This LED activity is useful for
indicating that the device is functioning properly.
Summary of Contents for TORKEL 910
Page 2: ......
Page 13: ...CS033663DE ZP CS01E TORKEL 900 series 13 4 Menu system...
Page 27: ...CS033663DE ZP CS01E TORKEL 900 series 27 6 External current measurement...
Page 31: ...CS033663DE ZP CS01E TORKEL 900 series 31 7 Alarm function...
Page 33: ...CS033663DE ZP CS01E TORKEL 900 series 33 8 How to obtain the desired current...
Page 41: ...CS033663DE ZP CS01E TORKEL 900 series 41 10 Troubleshooting...
Page 43: ...CS033663DE ZP CS01E TORKEL 900 series 43 11 Calibration...
Page 45: ...CS033663DE ZP CS01E TORKEL 900 series 45 12 Specifications...
Page 49: ...CS033663DE ZP CS01E TORKEL 900 series 49 12 Specifications...
Page 51: ......