ASP-I-MASPI-9 MedRx AVANT REMsp Installation Manual 

                                                                                                                                                                   Effective Date: 10-21-2016

    

12 

Electrical fast transient / 
burst  

IEC 61000-4-4 

+/- 2 kV for power supply 
lines 

+/- 1 kV for input / output 
lines 

+/- 2 kV for power supply 
lines 

+/- 1 kV for input / output 
lines 

Mains power quality should 
be that of a typical 
commercial or hospital 
environment. 

Power frequency (50/60 Hz)  

Magnetic field 

 

IEC 61000-4-8 

3 A/m 

3 A/m 

Power frequency magnetic 
fields should be at levels 
characteristic of a typical 
location in a typical 
commercial or hospital 
environment. 

 

 

Guidance and manufacturer’s declaration – electromagnetic immunity

 

The Avant REMsp is intended for use in electromagnetic environment specific below. The customer or the user of the 
Avant REMsp should assure that it is used in such an environment.

 

Immunity test 

IEC 60601- test level 

Compliance level 

Electromagnetic environment - 

guidance 

 

 

 

Portable and mobile RF 
communications equipment should be 
used no closer to any part of the Avant 
REMsp, including cables, than the 
recommended separation distance 
calculated from the equation applicable 
to the frequency of the transmitter. 

Recommended separation distance: 

Conducted RF   
 
IEC 61000-4-6 

 
3 V

eff

  

 
3 V

eff

 

 

d = 1,17 

P

×

   

Radiated RF  

 

IEC 61000-4-3 

 

3 V/m 

80 MHz to 2,5 GHz  

 

3 V/m 

 

 

d = 1,17 

P

×

  80 to 800 MHz 

 

d = 2,33 

P

×

  800 MHz to 2,5 GHz 

 

 

 

Where P is the maximum output power 
rating of the transmitter in watts (W) 
according to the transmitter 
manufacturer and d is the 
recommended separation distance in 
meters (m).  

Field strengths from fixed RF 
transmitters, as determined by an 
electromagnetic site survey 

a

, should be 

less than the compliance level in each 
frequency range 

b

Interference may occur in the vicinity of 
equipment marked with the following 
symbol: 

 

 

NOTE 1     At 80 MHz and 800 MHz, the higher frequency ranges applies. 

NOTE 2     These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption  

and reflection from structures, objects and people.  

Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land 
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically 
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site 
survey should be considered. If the measured field strength in the location in which the Avant REMsp is used 
exceeds the applicable RF compliance level above, the Avant REMsp should be observed to verify normal 
operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or 
relocating the Avant REMsp 

Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m. 

Summary of Contents for Avant REMsp

Page 1: ...INSTALLATION MANUAL www medrx usa com...

Page 2: ...ation 5 Driver Installation 8 EMC Precautions 11 Safety 14 Cleaning 15 Limited Warranty 16 0086 www medrx usa com 1200 Starkey Rd 105 Largo FL 33771 U S A Toll Free 888 392 1234 727 584 9600 Fax 727 5...

Page 3: ...of precision in situ verification for your office Compact yet rugged this PC Based system is USB powered and supports ANSI and IEC Real Ear tests The software includes targets for DSL i o 5 0 and NAL...

Page 4: ...19 39 from the patient for normal use It is preferred to plug the speaker into the power supply but it may be powered by the internal battery The Bluetooth option is not used The speaker s power cable...

Page 5: ...arts 2 On the Setup screen choose Install Avant REM If the Setup program does not automatically start Press the Win R keys on the keyboard Type D Setup exe in the window where D is the drive letter as...

Page 6: ...important document defines the acceptable usage of the REM software After reading the Agreement click I accept Click Next 6 This screen indicates the location of the program files Click Next Note You...

Page 7: ...AVANT REMsp Installation Manual Effective Date 10 21 2016 7 8 Ready to Install Program Click Install 9 Wait while the InstallShield Wizard installs the Avant REM program 10 When the installation is co...

Page 8: ...commend utilizing the supplied 6 inch or 6 foot USB extension cable with your REMsp Plug the USB extension cable into your computer as shown above 2 Wait for the system to copy and install default dri...

Page 9: ...anced Options screen click Install Drivers 6 When this screen appears click Ok 7 If this window appears and all devices are Ready to use click Close Drivers installed successfully 8 If this warning sc...

Page 10: ...en the audio properties are configured properly during driver installation the Audio Tab will appear like the image on the left If not use the pull down lists to adjust the settings to match the image...

Page 11: ...eclaration electromagnetic emissions The Avant REMsp is intended for use in electromagnetic environment specific below The customer or the user of the Avant REMsp should assure that it is used in such...

Page 12: ...P 80 to 800 MHz d 2 33 P 800 MHz to 2 5 GHz Where P is the maximum output power rating of the transmitter in watts W according to the transmitter manufacturer and d is the recommended separation dista...

Page 13: ...ut power of transmitter W Separation distance according to frequency of transmitter meters 150 kHz to 80 MHz d 1 17 P 80 MHz to 800 MHz d 1 17 P 800 MHz to 2 5 GHz d 2 33 P 0 01 0 12 0 12 0 233 0 1 0...

Page 14: ...Storage temperature range at least from 0 C to 50 C All components with patient contact are made of bio compatible materials This device does not produce any adverse physiological effects Install the...

Page 15: ...not be re used by another patient 2 It is recommended that 70 Isopropyl Alcohol should be applied to a soft clean cloth or tissue not directly on the component to be cleaned The cloth should never be...

Page 16: ...greater than 100 F and water liquid damage Repair or replacement of the system as provided under this warranty is the sole and exclusive remedy of the purchaser MedRx shall not be liable for any cons...

Reviews: