48
EVENCARE®
ProView
™
The temperature is above
or below the proper range
for testing.
The temperature should be 50˚F-104˚F.
Move the meter and test strips to an
area that is within the temperature range
to perform the test. Wait about twenty
minutes or until the meter has reached
the proper range and retest. Do not
raise or lower the meter temperature
artificially.
There is not enough blood
in the strip to perform a
test.
The wrong strip has been
put into the meter.
Repeat the test with a new test strip
and make sure enough blood has been
applied to the strip.
Make sure you are using an ProView test
strip and repeat the test.
What you see
What it means
What you should do
Troubleshooting