Disinfection Instruction:
Step 1. Before disinfecting, clean the meter as described in Cleaning Your Meter process.
Step 2. Wash hands with soap and water, put on single-use medical protective gloves.
Step 3. Prepare the CaviWipes towelette or other EPA-registered disinfecting wipe.
Take out a wipe from the container and follow the instructions on the package.
If needed squeeze the wipe slightly to remove the excess liquid.
Step 4. Wipe the glucose meter thoroughly including the front, back and sides, and take
care not to get any liquid in the test strip port and serial port. Do not wrap the
meter in a wipe.
Step 5. If using the CaviWipes towelette, allow to remain wet for two minutes. For other
EPA-registered disinfecting wipes, allow the surface of the meter to remain wet for
the contact time listed on the disinfecting wipe’s instructions for use. Dispose of
wipe when finished.
Step 6. After disinfection, user should take off gloves and wash hands thoroughly with
soap and water before proceeding to the next patient.
The EVENCARE ProView Blood Glucose Monitoring System was demonstrated that there
was no change in performance or in the external materials of the meter after 11,000
cleaning with disinfection cycles with the CaviWipes towelette designed to simulate three
years of device use.