23
EVENCARE®
ProView
™
Meter Setup
•
Record the date on the bottle when you open a new bottle of test strips. Discard
any unused test strips three months after opening.
•
Test strips are good three months after opening or until the last day of the month of
expiration, whichever comes first.
•
Make sure you are performing the test in an environment that is between
50˚F–104˚F.
Step 2. When the blood drop icon
mem
DAY
AVG
appears on the screen, shake the bottle of control
solution and then squeeze a drop of the control solution onto a clean, dry, non-
absorbent surface such as table top. Bring the test strip gently to the drop of
control solution so the sampling end of the test strip touches the control solution.
The drop of control solution will be drawn in automatically by the test strip.
IMPORTANT:
•
Replace the cap of the control solution immediately after use.
•
If there is insufficient sample, the strip icon will flash and the message “Er 4” will be
shown. Eject and discard the used test strip. Repeat the control solution test with a
new test strip.