Meter Setup
•
Record the date on the bottle when opening a new bottle of control solution.
Discard any unused control solution three months after the opening date.
•
Control solutions are good three months after opening date or until the last day
of the month of expiration, whichever comes first.
•
DO NOT FREEZE. Store the control solutions at room temperature of 39.2˚F – 86˚F.
Step 1. Insert a test strip into the meter with the arrow facing up.
Make sure to firmly and completely insert the test strip into
the meter. The meter will turn on automatically.
IMPORTANT:
•
Do not use test strips that are expired. Check the expiration date printed on the test
strip package.
•
Use each test strip immediately after removing it from the bottle.
•
Close the cap of the bottle immediately after removing a test strip.
•
Do not use wet, bent, scratched, or visibly damaged test strips.
•
Keep the test strips away from direct sunlight and heat. Store the test strip bottle in
a dry, cool place.